User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HEA 4934A/4935A HOLD TONE DROPOUT DETECTOR DATE: 20-Jul-93 AUTHOR: User Contributed REVISION: 1 F. Samonski 18-May-93 Added 4934A ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 8 NUMBER OF LINES: 88 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 HEAD [27][91]1m {TERMINATION IMPEDENCE CALIBRATION} 1.003 DISP Set the UUT Controls as follows: 1.003 DISP 1.003 DISP DISPLAY RCV 1.003 DISP REV ON 1.003 DISP Set up RCV 600 ohm ON 1.003 DISP MEASUREMENT NOISE WITH TONE 1.003 DISP FILTER C-MESSAGE 1.003 DISP 1.004 MESS Decrease the Output until ERR 7 appears on the UUT. 1.005 5700 10mV 5U 1004H N 2W 1.006 MESS 1.007 DISP 1.007 DISP Set the UUT to LEVEL FREQUENCY. 1.007 DISP 1.008 5700 mV 1004H S 2W 1.009 EVAL Does the UUT indicate between -44 and -48 dBm. 2.001 5700 * S 2.002 DISP 2.002 DISP REMOVE ALL CONNECTIONS FROM THE UUT. 2.002 DISP 2.003 OPBR Is the TI a 4934A ? 2.004 JMPT 5.002 2.005 DISP Set the UUT Controls as follows: 2.005 DISP 2.005 DISP MEASUREMENT LEVEL FREQUENCY 2.005 DISP REV & NOR BOTH ON 2.005 DISP SETUP BOTH 600 ON 2.005 DISP DISPLAY RCV 2.005 DISP BRG OFF 2.005 DISP set OUTPUT LEVEL for 0 dBm INDICATION. 2.006 DISP 2.006 DISP Set the BRG function to ON. 2.006 DISP 2.007 EVAL Does the UUT indicate between 6.0 and 6.2 dBm. 3.001 DISP 3.001 DISP Turn the BRG function off. 3.001 DISP 3.002 DISP 3.002 DISP Set both 135 ohm switches to ON. 3.002 DISP 3.003 EVAL Does the UUT indicate 0 dBm PLUS or MINUS 1 dBm. 4.001 DISP 4.001 DISP Set both 900 ohm switches to ON. 4.001 DISP 4.002 EVAL Does the UUT indicate 0 dBm PLUS or MINUS 1 dBm ? 5.001 JMP 8.001 5.002 DISP Set the UUT Controls as follows: 5.002 DISP 5.002 DISP MEASUREMENT LEVEL FREQUENCY 5.002 DISP REV OFF 5.002 DISP SETUP BOTH 600 ON 5.002 DISP DISPLAY RCV 5.002 DISP BRG OFF 5.002 DISP set OUTPUT LEVEL for 0 dBm INDICATION. 5.003 DISP 5.003 DISP Short the RCV and TRMT connectors together 5.004 DISP 5.004 DISP Set the BRG function to ON. 5.004 DISP 5.005 EVAL Does the UUT indicate between 6.0 and 6.2 dBm. 6.001 DISP 6.001 DISP Turn the BRG function off. 6.001 DISP 6.002 DISP 6.002 DISP Set both 135 ohm switches to ON. 6.002 DISP 6.003 EVAL Does the UUT indicate 0 dBm PLUS or MINUS 1 dBm. 7.001 DISP 7.001 DISP Set both 900 ohm switches to ON. 7.001 DISP 7.002 EVAL Does the UUT indicate 0 dBm PLUS or MINUS 1 dBm ? 8.001 END